🫀 ECG lead placement
10 electrodes. 12 views of the heart.
Use landmarks, not guesswork. Explain the test, obtain consent and offer a chaperone. Position comfortably, prepare the skin and keep the patient still.
Chest electrodes · IEC colours
| V1 | 4th intercostal space · right sternal edge. |
|---|---|
| V2 | 4th intercostal space · left sternal edge. |
| V4 | 5th intercostal space · left midclavicular line. |
| V3 | Halfway between V2 and V4. Place V4 first. |
| V5 | Left anterior axillary line · level with V4. |
| V6 | Left midaxillary line · level with V4 and V5. |
Find the sternal angle → 2nd rib → count spaces down. Place chest electrodes on the chest wall, beneath breast tissue when needed.
Limb electrodes · IEC colours
RA · right arm
Red
Red
LA · left arm
Yellow
Yellow
RL · right leg
Black
Black
LL · left leg
Green
Green
Use the limbs for a standard resting ECG. Follow cable labels: colour systems differ. Record any modified positions.
Common traps
- V1/V2 too high.
- V5/V6 following the rib down instead of staying level with V4.
- Arm leads reversed or limb electrodes moved onto the torso without recording it.